Original video: https://youtu.be/J7an1RLUJ8M
KURSUS UNTUK RAKYAT : COURSES FOR PEOPLE
O2 max is not ok, can ignore it because using 3rd party CO2 cartridge
Before taking measurements with the LI-6800 photosynthesis system, a warm-up is crucial for optimal performance. This involves running a series of automated tests that assess the functionality of key components like sensors, gas analyzers, and the environmental control system. The warm-up process ensures accurate readings by stabilizing the instrument's internal environment and checking for potential errors, ultimately increasing the reliability of the photosynthesis data collected.
Timestamp:
00:00 Warm-up test screen
00:13 CO2 test Not OK - Forgot to change for a new CO2 Cartridge
00:24 Warm up screen will tell warnings and failures during the tests, then suggest what should be done for the failed ones
00:35 Four chemical columns at the back of the machine console
00:40 CO2 mixer, Drierite (desiccant), Stuttgarter Masse (Ceramic Humidifier) and Soda lime (CO2 scrubber)
00:58 Change new 8gm CO2 Cartridge correctly
02:00 Redo warm-up test only for CO2 Limits and Match
02:55 Can use 3rd party CO2 cartridge food grade with
04:15 Result of warm-up if C